Table 8. Recommended separation distances between portable and m obile R F
communications equipment, and the PRM / ZWT (continued)
Separation distance according t o frequency of transmitter
a
m
Rated maximum
output power of
transmitter
bc
W
150 kHz to 80 MHz
d=1.2√P
80 MHz to 800
MHz
d
d=1.2√P
800 MHz to 2.5
GHz
d=2.3√P
10 3.8 3.8 7.3
100 12 12 23
a. The PRM and ZWT are intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The cu
stomer or the user can help prevent electromagnetic
interference by maintaining a minimum d istance between portable and mobile R F communications
equipment (transmitters), the PRM, and ZWT as recommended in the table, according to the
maximum output power of the communi
cations equipment.
b. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, a nd people.
c. For transmitters rated at a maximum o
utput power not listed in the table, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmit
ter manufacturer.
d. At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Specifications
Table 9. PRM Nominal Specifi cation
s
Characteristic
Nominal
Safety classification PRM: Class I.
ECG conn ection: Type BF,
defibrillation-protected.
Telemetry-wand connection: Type C
F,
defibrillation-protected
Ingress protection rating: IPX0
Dimensions
47 cm (18.5 in) deep, 36.8 cm (14.5 in
)
wide, 12.1 cm (4.75 in) high
Weight (app roxima te) 9.8 kg (2 1.5 lb)
Power rating 100–120 V 60 Hz, 220–240 V 50 Hz,
3.8/1.9 A
Power cord
2.4 m (8 ft), 100–240 V. Reliable
grounding is achieved only when
equipment is connected to receptacle
marked with “Hospital only” or “Hospital
grade.”
Duty cycle
Continuous
Operating temperature 10°C to 35°C ( 50°F to 95°F)
Transport and storage temperature
-40°C to 70°C (-40°F to 158°F)
Operating humidity 25% to 90%
40